• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX EPIDURAL MINIPACKS SYSTEM 1; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX EPIDURAL MINIPACKS SYSTEM 1; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number 100/391/116CZ
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/23/2021
Event Type  malfunction  
Manufacturer Narrative
A device history record (dhr) review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.The device was returned for evaluation.Visual and functional testing were performed.The connector has broken hinge.This reported problem has been raised to the supplier who recommended a redesign of the hinge area.Complaint trend was escalated to business unit level and the issue is being monitored.Mdr was generated under protocol (b)(4), as a result of warning letter cms# 617147.
 
Event Description
It was reported that the device broke.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX EPIDURAL MINIPACKS SYSTEM 1
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
olomoucka 306
hranice 1, mesto 753 0 1
EZ  753 01
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
olomoucka 306
hranice 1, mesto 753 0 1
EZ   753 01
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15326844
MDR Text Key305284649
Report Number3012307300-2022-16945
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K781052
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number100/391/116CZ
Device Lot Number4067835
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/11/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-