ST PAUL BCI SPECTRO2 PULSE OXIMETER Back to Search Results
Catalog Number 3301
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the device is showing error for mother board issue.No patient injury was reported.
 
Manufacturer Narrative
The returned device first time in for service over 15 years in the field and the failure resulted from corrupted eeprom memory chip due to wear of age.The device history record (dhr) review is not applicable or relevant because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.No information is provided to date.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# 617147.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI SPECTRO2 PULSE OXIMETER
Type of Device
OXIMETER
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15327103
MDR Text Key305385192
Report Number3012307300-2022-16980
Device Sequence Number1
Product Code DQA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number3301
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2020
Was the Report Sent to FDA? No
Date Manufacturer Received04/14/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-