MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number 37612 |
Device Problems
Electromagnetic Interference (1194); Failure to Interrogate (1332); Battery Problem (2885); Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that the healthcare provider (hcp) said they can't interrogate the patient's implant.The patient claimed that the airport scanners caused the implant to deplete per clinician.Technical services (ts) told caller that scanners are not likely the source of battery drain.The caller said the patient charged for 10 minutes, while caller was on the phone with ts, but patient stopped charging by the time the caller got back to him.Ts reviewed with caller that it takes about 4 hours to recharge fully.Hcp to let patient charge up further before attempting another interrogation.No symptoms were reported.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Search Alerts/Recalls
|
|
|