The product investigation was completed.Device evaluation details: visual analysis of the returned sample revealed a hole was found in the pebax, and the tip was found bent; no other damage or anomalies were observed on the smart touch unidirectional catheter.An irrigation test was performed, per bwi procedures.The catheter was irrigating correctly, no water leakage was observed during the investigation.The instructions for use contain the following recommendations: before use, check irrigation ports are patent by infusing heparinized normal saline through the catheter and tubing.Also, do not continue to use the catheter if still occluded or if it is not functioning properly.The event described could not be confirmed as the device performed without any irrigation issues.Although no irrigation defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the analysis.A manufacturing record evaluation was performed for the finished device (b)(4) number, and no internal actions related to the complaint were found during the review.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|