• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL BCI CAPNOCHECK II, MODEL 8400; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL BCI CAPNOCHECK II, MODEL 8400; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE Back to Search Results
Model Number 8401
Device Problems Device Alarm System (1012); Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection found no damage.Functional testing found occlusion error was present after powering up the monitor.Opened the monitor and found the tubing going to the bench was kinked.The root cause of the reported issue was found to be a known design issue where the weight of the restrictor pushes the small diameter tubing against the co2 bench nozzle as the housing closes.Tubing kit was replaced and monitor passed all functional testing.Operator of device is unknown.Initial reporter also sent report to fda is unknown.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# (b)(4).
 
Event Description
It was reported that the monitor reads continuous occlusion and was unable to obtain end tidal reading.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI CAPNOCHECK II, MODEL 8400
Type of Device
ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
6430 post rd
minneapolis, MN 55442
MDR Report Key15334166
MDR Text Key305594110
Report Number3012307300-2022-17156
Device Sequence Number1
Product Code CCK
UDI-Device Identifier10610586036781
UDI-Public10610586036781
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K991086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8401
Device Catalogue Number8401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/23/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-