Model Number CD3371-40QC
Device Problems Environmental Compatibility Problem (2929); Inappropriate or Unexpected Reset (2959); Compatibility Problem (2960)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/19/2022
Event Type  malfunction  
Manufacturer Narrative
Further information was requested but not received.
 
Event Description
It was reported, following an mri, the device was found to be in back up mode.Technical support was contacted and recommended reprogramming to resolve the event.The device was reprogrammed.The patient was stable.
 
Event Description
Additional information was received indicating the device was not programmed into mri mode prior to the patient undergoing and mri.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15334960
MDR Text Key305364549
Report Number2017865-2022-22076
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2022
Device Model NumberCD3371-40QC
Device Lot NumberP000099064
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/19/2022
Initial Date FDA Received09/01/2022
Supplement Dates Manufacturer Received09/06/2022
Supplement Dates FDA Received09/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/18/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial