Model Number CD3371-40QC
Device Problem Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/18/2022
Event Type  malfunction  
Event Description
It was reported that during generator replacement procedure, physician was unable to unscrew the screw holding the atrial lead in its port.Eventually, the existing atrial lead was capped and a new lead implanted.The generator change was completed.Patient was stable.
 
Manufacturer Narrative
The reported event of set screw issue was verified in the lab.Septum material was observed inside the is-1 set screw hex cavity and the set screw was found to be stripped.The connection issue was consistent with having occurred during the procedure.The device was above elective replacement indicator (eri) when received.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15339099
MDR Text Key304959380
Report Number2017865-2022-22118
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2017
Device Model NumberCD3371-40QC
Device Lot NumberA000004306
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/03/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-