MAUDE Adverse Event Report: ABBOTT VASCULAR EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM Back to Search Results
Model Number 22438-19
Device Problems Retraction Problem (1536); Deformation Due to Compressive Stress (2889)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/15/2022
Event Type  malfunction  
Event Description
It was reported that the procedure was to treat a lesion located in the internal carotid artery that was moderately calcified, heavily tortuous and 75% stenosed.At the end of the procedure and during attempted retrieval of the nav6 embolic protection device, resistance was felt when pulling back the barewire filter delivery wire to the retrieval catheter and it was observed that the guide wire was kinked in the vessel.Due to the resistance and kink, the guiding catheter was used to retrieve the nav6 and complete the procedure.There was no adverse patient effect and no clinically significant delay in the procedure.
 
Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow-up will be submitted with all relevant information.
 
Manufacturer Narrative
Visual was performed on the returned product.The retraction problem was unable to be confirmed due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.The investigation determined that the reported retraction problem was likely related to circumstances of the procedure.It is likely that barewire became kinked during use resulting in resistance while attempting to retract the filter into the retrieval catheter.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM
Type of Device
EMBOLIC PROTECTION SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15340489
MDR Text Key304533926
Report Number2024168-2022-09330
Device Sequence Number1
Product Code NTE
UDI-Device Identifier08717648137754
UDI-Public08717648137754
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K141678
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number22438-19
Device Catalogue Number22438-19
Device Lot Number2032361
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/15/2022
Initial Date FDA Received09/01/2022
Supplement Dates Manufacturer Received09/23/2022
Supplement Dates FDA Received10/13/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/23/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-