• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2 APP; DATA MANAGEMENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2 APP; DATA MANAGEMENT SYSTEM Back to Search Results
Model Number 71732-01
Device Problems Image Display Error/Artifact (1304); Unable to Obtain Readings (1516)
Patient Problems Hyperglycemia (1905); Loss of consciousness (2418)
Event Date 08/17/2022
Event Type  Injury  
Event Description
A complaint was received regarding an "unexpected application error" message from freestyle libre 2 application.As a result of not being able to log into the app to scan the sensor, on (b)(6) 2022, the customer had a loss of consciousness and was provided insulin by a non-healthcare professional for treatment.No further information was provided.There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
The product data has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Manufacturer Narrative
The reported complaint is an account password reset issue with the freestyle libre 2 version 2.7.2 app.The user received a message "unexpected application error" when pressing the "forgot password" link.An investigation was performed and determined that this was due to r8 obfuscation (the default compiler that converts java bytecode into the dex format that runs on the android platform.Obfuscation is used to reduce the size of the application by shortening the name of the classes and members¿ attributes), which had a conflict with the network library ¿retrofit¿.This caused an issue with the network call used to reset the password.Therefore, the complaint is confirmed.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A complaint was received regarding an "unexpected application error" message from freestyle libre 2 application.(b)(6) 2022, the customer had a loss of consciousness and was provided insulin by a non-healthcare professional for treatment.No further information was provided.There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2 APP
Type of Device
DATA MANAGEMENT SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key15340602
MDR Text Key299142148
Report Number2954323-2022-32064
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71732-01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/17/2022
Initial Date FDA Received09/01/2022
Supplement Dates Manufacturer Received10/10/2022
Supplement Dates FDA Received11/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/17/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-