It was reported that the procedure was to treat a lesion in the left carotid artery.The large emboshield nav6 embolic protection system (epd) was successfully positioned beyond the lesion and the lesion successfully stented; however, during removal, the epd became caught on a strut of the implanted stent, tearing the filter element.The epd remained in one piece and was able to be removed intact without damaging the implanted stent.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
|
Visual analysis were performed on the returned device.The reported torn filter element membrane was confirmed.The reported difficulty to remove from anatomy could not be confirmed as the exact conditions encountered by the device during the procedure could not be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported no similar incidents reported for this lot.The difficulty removing and the damage to the filter membrane was reportedly due to interaction with the implanted stent.Difficulty removing the filtration element from the delivery catheter pod, entanglement of filter element with stent, torn material can be affected by numerous factors including, but not limited to, use error, stent is placed too close to filter, excessive embolic load.The investigation was unable to determine a conclusive cause for the reported difficulties.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
|