The reported complaint of "all leds on the platform and the lcd backlight lit up, but the platform made a clicking noise and the lcd screen was blank" was not confirmed during testing at zoll.After the autopulse platform powered on, the lcd displayed was clear and readable (not blank), and the zoll service engineer did not report hearing any sort of clicking sound.During visual inspection, multiple cracks were observed on the top cover near the load plate cover.In addition, there was a vertical crack going through one of the screw fittings at the front-end area of the front enclosure, and there were multiple cracks in the screw well area of the bottom enclosure.The noted physical damages are unrelated to the reported complaint and are characteristics of harsh impacts due to user mishandling.All the damaged covers were replaced to address the issues.The customer was concerned if blood may have entered the platform during patient treatment at the hospital.The zoll service engineer removed both the front and bottom enclosures but did not observe any traces of blood ingress.The archive data review showed the occurrence of system error - 132 (internal watchdog timeout), unrelated to the reported complaint.The autopulse platform failed initial functional testing due to "system error, out of service, revert to manual cpr" displayed upon powering up, unrelated to the reported complaint.Technical investigation revealed that the root cause of the system error was the defective processor board, possibly due to a defective component.The processor board was replaced to remedy the fault.Following service, the autopulse platform was subjected to a run-in test using the 95% large resuscitation test fixture (lrtf) with known good test batteries until discharged without any fault or error.
|
The customer reported that the patient did not meet the indications for autopulse use, but the paramedic placed the patient on the autopulse platform (s/n (b)(4) proactively.The paramedic did not have to use the platform during transport to the hospital because the patient did not experience cardiac arrest.The crew members stated that blood may have entered the platform during patient treatment at the hospital.After returning to the station, the autopulse platform did not function as intended.When the autopulse was powered on, all leds on the platform and the lcd backlight lit up, but the platform made a clicking noise and the lcd screen was blank.No consequences or impacts on the patient.
|