MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK NEUROFORM ATLAS 4.0MM X 15MM NO TIP - CE; STENT, INTRACRANIAL NEUROVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK NEUROFORM ATLAS 4.0MM X 15MM NO TIP - CE; STENT, INTRACRANIAL NEUROVASCULAR Back to Search Results
Catalog Number M003EZAS40150
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/16/2022
Event Type  malfunction  
Manufacturer Narrative
The device is not available to the manufacturer.
 
Event Description
It was reported that during an endovascular procedure in a patient with a pcoma (posterior communicating artery) aneurysm, the deployed subject stent dislodged/migrated away from the target lesion and did not cover the neck of the aneurysm.The procedure was completed successfully by deploying another stent to cover the neck of the aneurysm.No clinical consequences were reported to the patient due to this event.
 
Manufacturer Narrative
The product was not returned for analysis therefore, visual and functional testing was not performed.Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.The reported event is covered in the device direction for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported event 'stent dislodged/migrated' could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Additional information provided by the customer was that after placing the stent to the lesion to deploy it, he stent suddenly retracted back and it resulted in the deployment of the stent which didn't cover the neck of aneurysm completely.It was also reported the device was prepared as per the dfu, that there was no damage noted to the packaging prior to opening and the device was confirmed to be in good condition during preparation/prior to use on the patient, and that continuous flush was set up and maintained during the procedure.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable will be assigned to this complaint.
 
Event Description
It was reported that during an endovascular procedure in a patient with a pcoma (posterior communicating artery) aneurysm, the deployed subject stent dislodged/migrated away from the target lesion and did not cover the neck of the aneurysm.The procedure was completed successfully by deploying another stent to cover the neck of the aneurysm.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROFORM ATLAS 4.0MM X 15MM NO TIP - CE
Type of Device
STENT, INTRACRANIAL NEUROVASCULAR
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15353508
MDR Text Key305493532
Report Number3008881809-2022-00436
Device Sequence Number1
Product Code NJE
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P180031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberM003EZAS40150
Device Lot Number22963586
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/16/2022
Initial Date FDA Received09/03/2022
Supplement Dates Manufacturer Received10/24/2022
Supplement Dates FDA Received11/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
6F ENVOY CATHETER (CODMAN NEURO); SL-10 MICROCATHETER (STRYKER); SYNCHRO GUIDEWIRE (STRYKER); TARGET 360 3*4 (STRYKER); TARGET HELICAL 1*2 (STRYKER); TARGET HELICAL 2*4 (STRYKER); TARGET 360 3*8 (STRYKER)
Patient Age67 YR
Patient SexMale
Patient Weight58 KG
Patient RaceAsian
-
-