SMITHS MEDICAL ASD, INC. CLEO 90 INFUSION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7230-24
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Injection Site Reaction (4562)
Event Date 12/03/2019
Event Type  malfunction  
Manufacturer Narrative
No device was returned.The following operations were reviewed, in order to verify that the operations were properly performed, also records were reviewed, in order to ensure that complied with gmp requirements and shm procedures: the 'assemble base with cannula to inserter' operation was reviewed to ensure that the cannula is properly assembled; no discrepancies were found.The operators that perform critical operations such as bond skin adhesive film onto flange of sleeve and the skin adhesive inspections using a monitor are certified.The operator performs the visual inspection of the adhesive using the skin adhesive inspection.4 samples from other lot number were taken from production floor and tested for in-process testing and no discrepancies were found.Visual inspection of 32 samples from a different lot number were taken from production line and were visually inspected in order to verify that the spring is not released.Results: no discrepancies were detected.The retractor assemblies were 100% visually inspected by the production personnel to assure that the needle is properly threaded.Production personnel verified 100% that parts were not damaged (including scratches) or distorted/warped.Quality personnel verifies 100% that the needle tip is up, and the blunt end is fully seated and verify the length while still in the needle hub on the pallet.No root cause could be determined, as no samples or pictures were received for evaluation.No corrective actions are required since the complaint could not be confirmed.A device history record (dhr) review could not be performed, since the lot number was not specified by the customer.It was reported that the patient experienced a reaction at the infusion site, described as edema.No medical or surgical intervention was reported.D5- operator of device patient/consumer this remediation mdr was generated under protocol b10009406, as a result of warning letter cms# 617147.
 
Event Description
It was reported that the patient experienced a reaction at the infusion site, described as edema.No medical or surgical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEO 90 INFUSION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15353767
MDR Text Key305708916
Report Number3012307300-2022-17606
Device Sequence Number1
Product Code FPA
UDI-Device Identifier30610586028391
UDI-Public30610586028391
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K042172
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number21-7230-24
Device Catalogue Number21-7230-24
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/21/2022
Initial Date FDA Received09/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-