OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS; RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS
|
Back to Search Results |
|
Model Number A42011A |
Device Problems
Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/01/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
The suspect medical device was returned to the manufacturer for investigation/evaluation.The investigation confirmed that the ceramic insulation at the distal end of the resection sheath is broken off.Based on the information provided about the event/incident the damage was most likely caused when the high-frequency current was activated while the distal end of the resectoscope/electrode was in contact with gas or located in its direct proximity.This caused an explosion due to the oxyhydrogen gas that is formed during the use of a resectoscope.Therefore, this event/incident was attributed to use error.Furthermore, the investigation found the laser marking to be faded, which was attributed to wear and tear.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the resection sheath without showing any abnormalities.The case will be closed on olympus side with no further actions.The reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
|
|
Event Description
|
Olympus was informed that during a thearpeutic transcervical resection (tcr) procedure, the ceramic insulation at the distal end of the resection sheath broke off, separating into many tiny fragments.A portable x-ray was used to locate the various fragments and remove them, which, however, caused a prolongation of the procedure of approx.30 minutes.
|
|
Search Alerts/Recalls
|
|
|