MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS; RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS; RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS Back to Search Results
Model Number A42011A
Device Problems Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/01/2022
Event Type  Injury  
Manufacturer Narrative
The suspect medical device was returned to the manufacturer for investigation/evaluation.The investigation confirmed that the ceramic insulation at the distal end of the resection sheath is broken off.Based on the information provided about the event/incident the damage was most likely caused when the high-frequency current was activated while the distal end of the resectoscope/electrode was in contact with gas or located in its direct proximity.This caused an explosion due to the oxyhydrogen gas that is formed during the use of a resectoscope.Therefore, this event/incident was attributed to use error.Furthermore, the investigation found the laser marking to be faded, which was attributed to wear and tear.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the resection sheath without showing any abnormalities.The case will be closed on olympus side with no further actions.The reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Event Description
Olympus was informed that during a thearpeutic transcervical resection (tcr) procedure, the ceramic insulation at the distal end of the resection sheath broke off, separating into many tiny fragments.A portable x-ray was used to locate the various fragments and remove them, which, however, caused a prolongation of the procedure of approx.30 minutes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS
Type of Device
RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM   22045
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662
MDR Report Key15354717
MDR Text Key299277481
Report Number9610773-2022-00366
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761023658
UDI-Public04042761023658
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K931994
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 09/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA42011A
Device Catalogue NumberA42011A
Device Lot Number19XW-0089
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/11/2022
Initial Date FDA Received09/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
OLYMPUS ESG-400 ELECTROSURGICAL GENERATOR
Patient SexFemale
-
-