• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JUNI OX FB FEM SZ 6 LM RL; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JUNI OX FB FEM SZ 6 LM RL; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71422346
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Insufficient Information (4580)
Event Date 05/07/2019
Event Type  Injury  
Event Description
It was reported that after a left tka performed on (b)(6) 2018 due to osteoarthritis, the patient experienced an unspecified adverse event.This was addressed via revision surgery on (b)(6) 2019 where all components where exchange for competitor´s implants.This information was provided by the national joint registry of the united kingdom, as part of a retrospective data collection of patients who underwent a revision surgery of journey uni prothesis.As such, no further information will be available.
 
Manufacturer Narrative
Internal complaint reference: (b)(6).
 
Manufacturer Narrative
The device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, a left revision was performed approximately 15 months post uka due to an unspecified adverse event.As of the date of this medical investigation, the requested clinical documentation has not been provided as the information was provided by the national joint registry of the united kingdom as part of a retrospective data collection, and reportedly, no further information will be available.Without the requested medical documentation, possible clinical factors which could have contributed to the unspecified adverse event could not be concluded.The patient impact beyond the reported revision could not be determined.No further medical assessment can be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for knee systems provides complete guidelines of indications, contraindications, warnings and precautions and possible adverse effects that may occur post operative.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.No details of the alleged fault, malfunction or injury were provided, therefore no probable cause can be delineated.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Internal complaint reference number: case-2022-00118652-1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JUNI OX FB FEM SZ 6 LM RL
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15358754
MDR Text Key299281185
Report Number1020279-2022-03970
Device Sequence Number1
Product Code HSX
UDI-Device Identifier03596010629777
UDI-Public03596010629777
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K073175
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Study,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/05/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71422346
Device Catalogue Number71422346
Device Lot Number17CM06525
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received09/16/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/08/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
JOURNEY UNI TIBINRT S5-6LM/RL8MM, LOT#:17BM00455.; JRNY UNI TIBIAL BASE LM/RL SZ 6, LOT#:16BM07969.
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age60 YR
Patient SexMale
-
-