As part of the investigation, instrumentation laboratory (il) reviewed the acl top family 50 series operator's manual.The investigation confirmed that there are appropriate instructions and cautions regarding the activities of a stuck or jammed cts piercer probe in the cap of a primary sample tube.Per the acl top family 50 series operator's manual; "warning piercing hazard: do not put hand inside instrument while sample arm is in motion.Caution biohazard: in the unlikely event that the cts piercer probe gets stuck or jammed in the cap of a primary sample tube, do not attempt to remove the cap.Call service for assistance.The cts piercer probe assembly is delicate, very sharp, and biohazardous.The removal of a stuck cap from the piercer probe could cause personal injury or damage to the instrument".The acl top 350 cts performed as intended with no malfunction and its labeling provides appropriate instructions and cautions to the user.Therefore, no remedial action is indicated.
|