Model Number 71922-01 |
Device Problems
Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
|
Patient Problems
Hypoglycemia (1912); Muscle Weakness (1967); Confusion/ Disorientation (2553)
|
Event Date 08/20/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A customer reported via webform that a sensor error message was received on the adc device.Due to this, the customer became hypoglycemic with symptoms of confusion and weakness and was unable to self-treat, requiring "fruit juices and 2 quick sugar" for treatment by a friend.No further information was reported.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.No product has been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the libre sensor and libre sensor kits were reviewed and the dhrs showed the libre sensor and libre sensor kits passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A customer reported via webform that a sensor error message was received on the adc device.Due to this, the customer became hypoglycemic with symptoms of confusion and weakness and was unable to self-treat, requiring "fruit juices and 2 quick sugar" for treatment by a friend.No further information was reported.There was no report of death or permanent injury associated with this event.
|
|