Model Number CLICKFINE 32GX4MM DONGBAO 7CT
Device Problems Loose or Intermittent Connection (1371); Mechanical Problem (1384)
Patient Problems Pain (1994); Dizziness (2194)
Event Date 07/18/2022
Event Type  Injury  
Manufacturer Narrative
The patient reported pain at the injection site and dizziness upon use of the clickfine needle.The patient reported that the clickfine could not be screwed tightly onto the pen which led to liquid leakage.The symptoms lessened after taking oral drugs and using a screw on brand needle.It was noted that this is the first initial uses of the patient of the clickfine needle after switching from a local brand.Clickfine is supposed to be clicked onto the pen (pushed in rather than screwed in) which might have contributed to the connection issue and leakage.The customer didn't return any needles for investigation.Other investigation included a review for any internal deviations during production, which showed no abnormalities or deviations from the validated manufacturing process.Penetration force testing was also performed with 5 samples retained from the same batch number.None of the needles tested from the same batch reached beyond the maximum penetration force of 1.0n (max 0.7 per gb15811-2016).
 
Event Description
Starting on the 5th of july, the patient switched from a local brand needle to clickfine needle for long-term insulin glargine injection.On the (b)(6) 2022, the patient went to the hospital and reported that the dongbao clickfine needle could not be screwed tightly on the pen and this lead to liquid leakage.The patient had experienced dizziness and pain at the injection site.The symptom lessened after replacing another brand needle and taking oral drugs.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLICKFINE 32GX4MM DONGBAO 7CT
Type of Device
PEN NEEDLE
Manufacturer (Section D)
YPSOMED AG
brunnmattstrasse 6
burgdorf, bern CH-34 01
SZ  CH-3401
Manufacturer (Section G)
YPSOMED AG
brunnmattstrasse 6
burgdorf, bern CH-34 01
SZ   CH-3401
Manufacturer Contact
clarisa tate
23 n echo park ct
mountain house, CA 95391
5105796072
MDR Report Key15366114
MDR Text Key299353778
Report Number3002806818-2022-00010
Device Sequence Number1
Product Code FMI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K171984
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/06/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberCLICKFINE 32GX4MM DONGBAO 7CT
Device Catalogue Number700000176
Device Lot Number201255-03
Was Device Available for Evaluation? No
Date Manufacturer Received08/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/25/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-