• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CORPORATION RX/5MM BASKET DIAMETER/180CM; TEMPORARY CAROTID CATHETER FOR EMBOLIC CAPTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION RX/5MM BASKET DIAMETER/180CM; TEMPORARY CAROTID CATHETER FOR EMBOLIC CAPTURE Back to Search Results
Model Number 501814RE
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/08/2022
Event Type  malfunction  
Manufacturer Narrative
A review of the manufacturing documentation associated with lot 35265330 presented no issues during the manufacturing process that can be related to the reported event.This device has been analyzed but the final, approved draft of the engineering report and its conclusion is not yet available.However, it will be submitted within 30 days upon receipt.Additional information is pending and will be submitted within 30 days upon receipt.
 
Event Description
As reported, the 5mm basket diameter/180cm angioguard rx emboli capture guidewire was unable to be inserted into the sheath due to the the angioguard filter, which was detached from the wire.Upon opening the package, it was noted that the deployment sheath tip was not fully seated in the filter basket introducer.It was already broken upon opening package.A 6.0mm angioguard was then used and the procedure was completed successfully.There were no reports of patient injury.The product was stored, handled, inspected and prepped according to the instructions for use (ifu) and was already broken prior to use.The device was intended for use in a carotid stenting procedure.The torque device locked onto the guidewire.The anti-migration clip located closest to the filter introducer was left in place until after the filter basket was docked into the deployment sheath.The proximal end of the deployment sheath was in contact with the torque nut prior to removing the device from the coil dispenser.The device will be returned for evaluation.
 
Manufacturer Narrative
Complaint conclusion: the 5mm basket diameter/180cm angioguard rx emboli capture guidewire was unable to be inserted into the sheath due to the angioguard filter which was detached from the wire.Upon opening the package, it was noted that the deployment sheath tip was not fully seated in the filter basket introducer.It was already broken upon opening package.A 6.0mm angioguard was then used and the procedure was completed successfully.There were no reports of patient injury.The product was stored, handled, inspected and prepped according to the instructions for use (ifu) and was already broken prior to use.The device was intended for use in a carotid stenting procedure.The torque device locked onto the guidewire.The anti-migration clip located closest to the filter introducer was left in place until after the filter basket was docked into the deployment sheath.The proximal end of the deployment sheath was in contact with the torque nut prior to removing the device from the coil dispenser.The product was returned for analysis.One non-sterile angioguard rx/5mm basket diameter/180cm unit was received coiled for analysis inside a plastic bag.The device was unpacked to proceed with the product evaluation.During visual inspection, the embolic capture guide wire with the torque device not affixed and the capture sheath were returned for evaluation.The filter basket was noted deployed.No other anomalies were observed on the components during the analysis.Functional analysis could not be performed, due to the condition of the unit as received, and since the deployment sheath was not returned for evaluation, only the ecgw which was already peeled off.A product history record (phr) review of lot 35265330 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿delivery system- loading (docking) difficulty - through introducer¿ was not confirmed through analysis of the returned device as the concomitant sheath was not returned and the condition of the returned device was not conducive for functional testing.The reported ¿ecgw (embolic capture guidewire) - fractured/separated - during prep¿ was not confirmed through analysis of the returned device as no fractures or separations were noted.The exact cause of the event could not be determined during analysis.Based on the information available for review, procedural or handling factors may have contributed to the event.According to the precautions in the safety information of the instructions for use, ¿guidewires are delicate instruments and should be handled carefully.Prior to use and when possible during the procedure, inspect the guidewire carefully for coil separation, bends, kinks, or damage of the filter basket assembly.The deployment and capture sheaths are delicate instruments and should be handled carefully.Prior to use, and when possible during the procedure, inspect the deployment sheath, capture sheath, and capture sheath rx port region (approximately 30 cm from the distal tip) for bends, kinks, or other damage.Do not use defective equipment.¿ neither the phr review nor the product analysis suggests that the event experienced by the customer could be related to the manufacturing process; therefore, no corrective/preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX/5MM BASKET DIAMETER/180CM
Type of Device
TEMPORARY CAROTID CATHETER FOR EMBOLIC CAPTURE
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60th ave
miami lakes FL 03246
Manufacturer Contact
karla castro
14201 nw 60th ave
miami lakes, FL 33014
7863138372
MDR Report Key15369385
MDR Text Key302938850
Report Number1016427-2022-05419
Device Sequence Number1
Product Code NTE
UDI-Device Identifier10705032054775
UDI-Public(01)10705032054775(17)240331(10)35265330
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K062531
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2024
Device Model Number501814RE
Device Catalogue Number501814RE
Device Lot Number35265330
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/12/2022
Initial Date FDA Received09/07/2022
Supplement Dates Manufacturer Received09/08/2022
Supplement Dates FDA Received09/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/06/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
6.0MM ANGIOGUARD, UNKNOWN SHEATH
-
-