Model Number 4542 |
Device Problems
Failure to Capture (1081); Fracture (1260); High impedance (1291)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/25/2022 |
Event Type
Injury
|
Event Description
|
It was reported that this left ventricular (lv) lead exhibited pace impedance measurements of greater then 3,000 ohms and loss of capture.It was noted that the lead was likely fractured.A lead revision procedure was performed in which the lv lead was explanted.A new lead was implanted and remains in service.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, visual inspection confirmed the lead was severed -12.5 cm from the terminal end.A facture of all conductors was observed -35 cm from the terminal end.
|
|
Event Description
|
It was reported that this left ventricular (lv) lead exhibited pace impedance measurements of greater then 3,000 ohms and loss of capture.It was noted that the lead was likely fractured.A lead revision procedure was performed in which the lv lead was explanted.A new lead was implanted and remains in service.No additional adverse patient effects were reported.
|
|