• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT IRELAND ALINITY I HBSAG QUALITATIVE II REAGENT KIT; TEST, HEPATITIS B (B CORE, BE ANTIGEN, BE ANTIBODY, B CORE IGM)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT IRELAND ALINITY I HBSAG QUALITATIVE II REAGENT KIT; TEST, HEPATITIS B (B CORE, BE ANTIGEN, BE ANTIBODY, B CORE IGM) Back to Search Results
Catalog Number 08P10-22
Device Problem False Positive Result (1227)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/20/2022
Event Type  malfunction  
Event Description
The customer stated that repeat reactive alinity i hbsag qualitative ii results were generated for a patient (b)(6).On (b)(6)2022 initial result 5.12 s/co (reactive); alinity i hbsag qualitative ii confirmatory 85% (confirmed positive).On (b)(6)2022 repeat result 1.4 s/co (reactive).The reactive results were not consistent with the patient's clinical history.A new sample (b)(6) was drawn from the same patient on (b)(6) and the result was nonreactive at 0.32 s/co.No impact to patient management was reported.
 
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
 
Manufacturer Narrative
Trending review determined no adverse trend for the issue for the product.Historical complaint review determined there is normal complaint activity for the lot number.Device history record review did not show any potential non-conformances, or deviations.Labeling was reviewed and found to adequately address the issue under review.Historical performance in the field of reagent lots using worldwide data through abbottlink was evaluated.The patient median result for the lot is comparable with all other lots in the field and within established baselines, confirming no systemic issue for the product lot.Based on the investigation, no deficiency for lot number 39377fn00 was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY I HBSAG QUALITATIVE II REAGENT KIT
Type of Device
TEST, HEPATITIS B (B CORE, BE ANTIGEN, BE ANTIBODY, B CORE IGM)
Manufacturer (Section D)
ABBOTT IRELAND
finisklin business park
sligo F91VY 44
EI  F91VY44
Manufacturer (Section G)
ABBOTT IRELAND
finisklin business park
sligo F91VY 44
EI   F91VY44
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key15374053
MDR Text Key300215132
Report Number3008344661-2022-00103
Device Sequence Number1
Product Code LOM
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P110029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/12/2023
Device Catalogue Number08P10-22
Device Lot Number39377FN00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/09/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ALNTY I HBS AG QUAL CON, 08P11-22, 39377FN00; ALNTY I PROCESSING MODU, 03R65-01, (B)(6); ALNTY I PROCESSING MODU, 03R65-01,(B)(6)
-
-