• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JUNI OX FB FEM SZ 4 RM LL; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JUNI OX FB FEM SZ 4 RM LL; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71422354
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Pain (1994)
Event Date 10/06/2015
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that, after a right uka surgery performed in (b)(6) 2013, the patient experienced unexplained pain.This event was addressed via revision surgery on (b)(6) 2015 to explant the femoral and tibial components.Patient health status is not known.This information was provided by the national joint registry of the united kingdom, as part of a retrospective data collection of patients who underwent a revision surgery of journey uni prothesis.As such, no further information will be available.
 
Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, a right revision was performed approximately 2 years post uka due to unexplained pain.The patient body mass index (bmi) is documented as 45; however, as of the date of this medical investigation, the requested clinical documentation has not been provided as the information was provided by the national joint registry of the united kingdom as part of a retrospective data collection, and reportedly, no further information will be available.Although the jii-uni surgical technique lists contraindications that tend to place increased loads on implants (such as age, weight, activity level, etc¿) it is unknown if the patient¿s bmi was a contributing factor to the reported event based on the limited information provided and it cannot be concluded that there was a malperformance of the component(s).The patient outcome beyond that which was reported could not be confirmed nor concluded; therefore, no further medical assessment can be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for knee systems revealed that improper fixation, and/or migration of the components could cause possible adverse effects.Besides, it establish that the implant can become damaged as a result of strenuous activity or trauma.The patient should be warned of surgical risks, and made aware of possible adverse effects, this has been identified as a warnings and precautions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, joint tightness, material in use, patient reaction or loss of sterility.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JUNI OX FB FEM SZ 4 RM LL
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15375167
MDR Text Key299444515
Report Number1020279-2022-04002
Device Sequence Number1
Product Code HSX
UDI-Device Identifier03596010627407
UDI-Public03596010627407
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K073175
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Study,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/07/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/06/2023
Device Model Number71422354
Device Catalogue Number71422354
Device Lot Number13DM05712
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received09/08/2022
Date Device Manufactured04/08/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
JRNYUNI APTIBSZ3RM/LL7MM, LOT#:10DM06250
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age59 YR
Patient SexFemale
Patient Weight110 KG
-
-