• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VDW GMBH VDW.SILVER RECIPROC; CONTROLLER, FOOT, HANDPIECE AND CORD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VDW GMBH VDW.SILVER RECIPROC; CONTROLLER, FOOT, HANDPIECE AND CORD Back to Search Results
Catalog Number V041163000000
Device Problems Mechanical Problem (1384); Energy Output Problem (1431)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Even if no serious injury resulted in this event, calibration related issues in endo motors have led to file separation in the past.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it is reported that vdw silver reciproc device having speed issue during use, speed of device is inconsistent.Outcome of injury is unknown as of this mdr and is pending.
 
Manufacturer Narrative
Battery (voltage collapses under load) defective, replaced.The housing was broken (probably due to a fall) and had an impact on the function (defective).When the housing was renewed, the display pane and the membrane keyboard were also replaced.Micromotor smr 118616, contra-angle 138004 (2019) and foot control 190417 checked, without errors.Sent without power supply.Function test without error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VDW.SILVER RECIPROC
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
VDW GMBH
bayerwaldstrasse 15
munich bayern 81737
GM  81737
Manufacturer (Section G)
VDW GMBH
bayerwaldstrasse 15
munich bayern 81737
GM   81737
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key15378799
MDR Text Key304356947
Report Number9611053-2022-00624
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 10/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Catalogue NumberV041163000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2022
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date09/06/2022
Initial Date Manufacturer Received 09/06/2022
Initial Date FDA Received09/08/2022
Supplement Dates Manufacturer Received09/06/2022
Supplement Dates FDA Received10/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage A
Patient Sequence Number1
-
-