• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS VASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS VASCULAR STENT GRAFT Back to Search Results
Catalog Number FVL10100
Device Problems Fracture (1260); Misfire (2532); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/22/2022
Event Type  malfunction  
Event Description
It was reported that during a stent graft placement procedure through femoral access via superficial femoral artery, the stent allegedly had partial deployment.The procedure was completed using another device.There was no reported patient injury.
 
Manufacturer Narrative
The catalog number identified has not been cleared in the us but is similar to the fluency plus endovascular stent graft products that are cleared in the us.The pro code and 510 k number for the fluency plus endovascular stent graft products are identified in common device name and pma/510k.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.Expiry date: 01/2025.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the fluency plus endovascular stent graft products that are cleared in the us.The pro code and 510 k number for the fluency plus endovascular stent graft products are identified in d2 and g4.H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment investigation summary: the physical sample was returned for evaluation.Based on the investigation performed in the laboratory, the stent graft was found partially deployed and the outer sheath found fractured.The metal guiding tube was found bent, probably due to the high release forces applied to the system.In this case, the system was flushed, and appropriate introducer sheath and guidewire were used.The vessel was not tortuous in nature or calcified.Additionally, it is known that the system was straightened prior to deployment, and that the proximal end of the stent graft was placed in a straight section of the lumen prior to deployment.The system was used to cover an aneurysm, which is an off-label use.Therefore, based on the investigation of the provided information, the investigation is confirmed for the partial deployment of the stent graft and subsequent fracture the outer sheath and deformation of the metal guiding tube.A definite root cause for the reported event could not be determined based upon the available information.Labeling review: in reviewing the relevant labeling, it was found that the instructions for use sufficiently address the potential risks.Regarding preparation of the device the instructions for use state that 'prior to loading the vascular system over a guide wire, both ports must be flushed with sterile saline (.).Flushing these lumens will also facilitate stent graft deployment.' regarding the anatomy of the placement site the instructions for use states: 'prior to stent graft deployment (.), ensure that the proximal stent graft end is positioned in a straight section of the lumen to reduce the risk of increased deployment forces and possible failure to deploy.' regarding accessories the instructions for use states: 'a super stiff guide wire (0.035 in.) is advanced from a femoral artery puncture site.Use an introducer sheath for the implant procedure'; the packaging pictograms indicate an introducer size of 9f and a 0.035" guidewire.It also contains the following precaution: 'the safety and effectiveness of the device for use in the treatment of aneurysms and pseudoaneurysms have not been established'.H10: d4 (expiry date: 01/2025), g3, h6 (device).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a stent graft placement procedure through femoral access via superficial femoral artery, the stent allegedly had partial deployment.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS VASCULAR STENT GRAFT
Type of Device
VASCULAR STENT GRAFT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15398840
MDR Text Key303333295
Report Number9681442-2022-00275
Device Sequence Number1
Product Code PFV
UDI-Device Identifier00801741145247
UDI-Public(01)00801741145247
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
P130029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberFVL10100
Device Lot NumberANGN2033
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/23/2022
Initial Date FDA Received09/12/2022
Supplement Dates Manufacturer Received10/07/2022
Supplement Dates FDA Received10/17/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age53 YR
Patient SexMale
-
-