• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VDW GMBH VDW.SILVER RECIPROC; CONTROLLER, FOOT, HANDPIECE AND CORD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VDW GMBH VDW.SILVER RECIPROC; CONTROLLER, FOOT, HANDPIECE AND CORD Back to Search Results
Catalog Number V041163000000
Device Problems Mechanical Problem (1384); Failure to Power Up (1476); Output Problem (3005)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
In this event it is reported that a vdw silver reciproc stops during use in treatment or does not turn on.Outcome of patient injury is pending.
 
Manufacturer Narrative
There has been a previous report received where this malfunction has caused file separation.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Manufacturer Narrative
Additional information was received indicating that no injury resulted.Battery (voltage collapses under load) defective, motor connection cable (torn strain relief) defective, foot control 103543 (loose contact) defective, micromotor smr16409 tested, without errors.Supplied without contra angle and power supply.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VDW.SILVER RECIPROC
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
VDW GMBH
bayerwaldstrasse 15
munich bayern 81737
GM  81737
Manufacturer (Section G)
VDW GMBH
bayerwaldstrasse 15
munich bayern 81737
GM   81737
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
MDR Report Key15400906
MDR Text Key305801614
Report Number9611053-2022-00625
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 10/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberV041163000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date09/09/2022
Initial Date Manufacturer Received 09/09/2022
Initial Date FDA Received09/12/2022
Supplement Dates Manufacturer Received09/09/2022
Supplement Dates FDA Received10/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage A
Patient Sequence Number1
-
-