• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MOMENTUM CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MOMENTUM CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR Back to Search Results
Model Number G125
Device Problems High impedance (1291); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/23/2022
Event Type  malfunction  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited high out-of-range shock impedance measurements on the non-boston scientific right ventricular (rv) lead.Setscrew issues were suspected.Additionally, the patient heard a beep tone from the device.Further evaluation was recommended.At this time, the product remains in service and no adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15407467
MDR Text Key304694364
Report Number2124215-2022-35577
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589126
UDI-Public00802526589126
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberG125
Device Catalogue NumberG125
Device Lot Number153695
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/23/2022
Initial Date FDA Received09/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/13/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age37 YR
Patient SexMale
-
-