DEPUY SPINE INC EXPEDIUM SFX CROSS CONNECTOR SYSTEM CONNECTOR A4 5.5 X 37-41MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 189401404
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/19/2022
Event Type  malfunction  
Event Description
Device report from synthes reports an event in south korea as follows: it was reported on august 19, 2022 that the center setscrew of this implant was not tightened when distributor opened its package for inventory entrance inspection.It seemed to be in the wrong position.When the distributor loosened the set screw, they found this thread to be worn out.This report involves one expedium sfx cross connector system connector a4 5.5 x 37-41mm.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional product codes: osh, mni, kwp, kwq, mnh.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: part # 189401404 lot # om11720 supplier: (b)(4) batch1: lot qty of (b)(4) unit were released on 02 mar 2022with no discrepancies.Batch2: lot qty of (b)(4) unit were released on 17 jan 2022 with no discrepancies.No ncrs were generated during production.The ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the head and threads of one of the set screw for sfx,5.5,ti, med, size a4 was observed stripped.The rest of the surface of the device shows normal wear which would not contribute to the complaint condition.A dimensional inspection for the sfx,5.5,ti, med, size a4 was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the sfx,5.5,ti, med, size a4 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPEDIUM SFX CROSS CONNECTOR SYSTEM CONNECTOR A4 5.5 X 37-41MM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15410959
MDR Text Key300231515
Report Number1526439-2022-01579
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034168111
UDI-Public10705034168111
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K142460
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number189401404
Device Catalogue Number189401404
Device Lot NumberOM11720
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/19/2022
Initial Date FDA Received09/13/2022
Supplement Dates Manufacturer Received11/01/2022
Supplement Dates FDA Received11/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/17/2022
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNKNOWN LOCKING/SET SCREWS
-
-