• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JOURNEY AP CUT BLOCK IMPACTOR; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JOURNEY AP CUT BLOCK IMPACTOR; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 74012421
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/23/2022
Event Type  malfunction  
Event Description
It was reported that, during set up and inspection for a tka surgery, it was found that a journey ap cut block impactor was missing both bearings and screws.As this was noticed before the procedure, the patient was not yet involved.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
Additional information: d4.
 
Manufacturer Narrative
The device was not returned for evaluation and the reported event could not be confirmed.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Manufacturer Narrative
This complaint has been re-evaluated for mdr reporting and it was determined that this case does not meet the threshold for reporting and is a non-reportable event, as the event happened in a not surgical enviroment; therefore, there was no patient involved.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly, and a further report submitted outlining both the event details and our investigations performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY AP CUT BLOCK IMPACTOR
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15411699
MDR Text Key303621572
Report Number1020279-2022-04071
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010556721
UDI-Public03596010556721
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 07/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74012421
Device Catalogue Number74012421
Device Lot Number14CMO5644
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/23/2022
Initial Date FDA Received09/13/2022
Supplement Dates Manufacturer Received09/06/2022
10/05/2022
07/19/2023
Supplement Dates FDA Received09/20/2022
10/06/2022
07/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/10/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-