Model Number EL5ML
Device Problem Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/19/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # v95u3l.Additional information additional information was requested and the following was obtained: "have tried finding out how the clips misfired but have not been able to get a response." investigation summary the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was received with no damage to the external components.In addition, the tyvek was returned along with the instrument.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, it was noted that the clips were not fed into the jaws in the next actuation of the trigger.In order to evaluate the condition of the internal components of the device, it was disassembled.Upon disassembling, the feed link was noted broken causing the feeding issues and 3 clips were found inside the clip track.The event reported was confirmed and it is related to improper use of the device.Possible causes for the damage found is due to the jaws may have been restricted during firing, or not fully through the trocar during firing.Please reference the instruction for use for more information.A manufacturing record evaluation was performed for the finished device lot and batch number, and no non-conformances were identified.
 
Event Description
It was reported that during a laparoscopic vaginal hysterectomy the device misfired.Continued with like product.No patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
329348013
MDR Report Key15412879
MDR Text Key306096798
Report Number3005075853-2022-06052
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberV95U3L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2022
Initial Date Manufacturer Received 08/18/2022
Initial Date FDA Received09/13/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/18/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-