EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER; CATHETER, FLOW DIRECTED Back to Search Results
Model Number D97120F5
Device Problem Pacing Problem (1439)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The pacing catheter will not be returned for evaluation, as it was discarded at the facility.Without the return of the unit, it is not possible to determine if some damage or defect existed on the unit that could have contributed to the event.It is not known if some procedural factors may have contributed to the event.A device history record review could not be completed because the lot number is unknown.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.Device discarded at facility.
 
Event Description
It was reported that during use, a physician was unable to get the pacing catheter (model d97120f5) to work.The pacer wire was inserted and turned on, but the device would not capture.It was removed and another device was used.The exact date of the event and the lot number are unknown.There was no patient injury.The product is not available for return as it was said to have been discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ BIPOLAR PACING CATHETER
Type of Device
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
jessica atallah
1 edwards way
irvine, CA 92614
9492500294
MDR Report Key15414724
MDR Text Key300177107
Report Number2015691-2022-07857
Device Sequence Number1
Product Code DYG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K822723
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberD97120F5
Device Catalogue NumberD97120F5
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/23/2022
Initial Date FDA Received09/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-