• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS LUCEA 40; DEVICE, MEDICAL EXAMINATION, AC POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS LUCEA 40; DEVICE, MEDICAL EXAMINATION, AC POWERED Back to Search Results
Model Number ARD568601997
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/07/2022
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.Device not returned to manufacturer.
 
Event Description
On 7th september, 2022 getinge became aware of an issue with one of examination lights - lucea 40.It was stated.The light cover was broken with missing particles due to physical damage.There was no injury reported, however, we decided to report the issue in abundance of caution as any detachment of part or particles may lead to potential infection of the patient.
 
Manufacturer Narrative
Getinge became aware of an issue with one of examination lights - lucea 40.It was stated the light cover was broken with missing particles due to impact damage.There was no injury reported, however, we decided to report the issue in abundance of caution as any detachment of part or particles may lead to potential infection of the patient.According to the information provided by getinge technician, the affected device was repaired by parts replacement and returned to use after passing functionality tests.Based on the information collected, it was established that when the event occurred, the examination light did not meet its specification, since cracks in covers, resulting in missing plastic particles could be considered as technical deficiency, and in this way the device contributed to the event.The provided information does not indicate if upon the event occurrence, the device was or was not being used for patient treatment.When reviewing reportable events for this type of issue we were able to establish that the received incidents are occurring at a moderate ratio.We have been able to confirm that the investigated issue has never led to serious injury or worse, to our knowledge.According to the analysis performed by the subject matter expert, the probable causes of the breakage are the incompatibility of the cleaning products or abnormal use.Based on information provided by getinge technician, it was confirmed that lighthead cover was broken due to impact damage.Maquet sas recommends visual inspection before use according to the user manual (ifu 01701 rev.11, page 21).A daily inspection performed by the user (chipped paint, impact marks and any other damage) will help preventing such event.Maquet sas recommends to inform the customers about the hazards in cases of non-compliance of these instructions.We believe the related devices are performing correctly in the market.We also believe that if the manufacturer¿s recommendation had been followed the incident could have been avoided.Getinge shall continue to monitor for any further events of this nature and do not propose any further action at this time.The correction of h4 device manufacture date field deems required.This is based on the internal evaluation.Previous h4 device manufacture date: 11/13/2019.Corrected h4 device manufacture date: 11/15/2019.
 
Event Description
Manufacturer's reference number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUCEA 40
Type of Device
DEVICE, MEDICAL EXAMINATION, AC POWERED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key15417486
MDR Text Key299868706
Report Number9710055-2022-00372
Device Sequence Number1
Product Code KZF
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberARD568601997
Device Catalogue NumberARD568601997
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/14/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/15/2019
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-