• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS LUCERA ELITE COLONOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS LUCERA ELITE COLONOVIDEOSCOPE Back to Search Results
Model Number PCF-H290DI
Device Problems Corroded (1131); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
A user facility submitted a repair request to the olympus service center, for the evis lucera elite colonovideoscope exhibiting e315 scope error.There was no report of patient harm or user injury associated with this event.
 
Manufacturer Narrative
The customer suspected device was returned for investigation.The olympus service center confirmed the reported event during inspection and testing.Upon evaluation of the returned device the following defects were found, corrosion of the scope connector causes the e315 scope error, curved rubber adhesive missing, bending angle insufficient due to elongation of the angle wire, play of the angle knob out of the normal state due to elongation of the angle wire, liquid leakage to the scope connector, scratches found on scope connector, scratches on insertion tube, dirt on insertion tube and operation part that¿s difficult to remove, scratches found on operation part, scratches found on tip cover, scratches found on switch box, scratches on operation unit cover, scratches on up/down knob, scratches on up/down angle fixing lever, scratches found on right/left knob, scratches found on the grip, scratches found on the universal cord, scratches on the scope connector side of the universal cord, scratches on scope connector cover, and scratches on the right/left angle fixing knob.The faulty parts were replaced, and the device was returned to the user facility.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the cause of the event is likely due to defective image sensor unit, or the failure of the internal circuit board of video connector or the system caused the event.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA ELITE COLONOVIDEOSCOPE
Type of Device
COLONOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15420009
MDR Text Key305889977
Report Number9610595-2022-01803
Device Sequence Number1
Product Code FDF
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPCF-H290DI
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/16/2022
Initial Date FDA Received09/14/2022
Supplement Dates Manufacturer Received10/17/2022
Supplement Dates FDA Received11/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/17/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-