NULL PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C37101324-NLJ |
Device Problem
Defective Device (2588)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/27/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that during the use of the product, the customer noticed a pinhole in the breathing bag.No patient injury was reported.
|
|
Manufacturer Narrative
|
A product sample and pictures were received for evaluation.Visual and functional testing were performed.A tear is observed in the pictures.The sample was received in used conditions without its original packaging.The sample was visually inspected at a distance of 12" under normal lighting to received unit, in order to detect any damaged components.The reported issue was confirmed.The breathing bag had a pinhole.Based on the sample provided, physically evaluation, the most probable cause is that the product became damaged after it left the manufacturing facility.No corrective actions are required since failure mode is not attributable at the manufacturing process.Udi is unknown.Operator of device is unknown.Manufacturing site address is unknown.Pma/510k is unknown.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
|