• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number UNKNOWN
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/19/2022
Event Type  malfunction  
Event Description
Livanova received report of scp control alarming during procedure.No further detail was provided and a reportable malfunction cannot be excluded.There was no patient injury.
 
Manufacturer Narrative
There was no patient involvement.Serial number is unknown.This information will be provided in a supplemental report if made available.As the serial number is unknown, the device manufacture date could not be determined.This information will be provided in a supplemental report if made available.Livanova deutschland manufactures the centrifugal pump system with tubing clamp.The incident occurred in new york, new york.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: through follow-up communication livanova clarified the reported issue and learned that the flow sensor triggered an alarm and the values were intermittent on the display.The reported failure does not stop or modify the pump flow.The flow rate can still be controlled with the knob and rpm can be estimated based on the leds around the control knob.Alarm system remain functional as well as any safety feature and alarms are displayed on system panel.Therefore the reported issue is not likely to result in serious injury and has been re-assessed as not reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key15428168
MDR Text Key300073264
Report Number9611109-2022-00468
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUNKNOWN
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-