Model Number CD3371-40C
Device Problems Premature Discharge of Battery (1057); Failure to Interrogate (1332); No Pacing (3268)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Further information was requested but not received.
 
Event Description
During an in clinic follow-up, the device was interrogated and was unable to communicate.The electrocardiogram (ecg) only showed the patients rhythm, and no device pacing was observed.Premature depletion of the battery was suspected.Device replacement is anticipated.The patient was in stable condition.
 
Manufacturer Narrative
Premature battery depletion, no pacing output and communication failure was confirmed by analysis.The loss of communication and no pacing output was due to low battery voltage; the low battery voltage was a result of premature battery depletion.A device evaluation was performed, and no sources of high current were noted.The cause of the premature battery depletion was consistent with lithium clusters formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Lithium clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15428961
MDR Text Key304371168
Report Number2017865-2022-22406
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2016
Device Model NumberCD3371-40C
Device Lot Number4437166
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0117-2017
Patient Sequence Number1
-
-