VENUSA DE MEXICO S.A. DE C.V. TACTICATH¿ QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
Model Number PN-004 075 |
Device Problems
Fluid/Blood Leak (1250); Electrical Shorting (2926)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/15/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
One 75 mm tacticath quartz contact force ablation catheter was received for evaluation.Multiple short circuits were detected.The device did not meet specifications during a shaft leak test and an irrigation leak test.The stainless steel irrigation tubing was fractured and detached from the deformable body at the location of the fractured spine, consistent with the failed leak tests, fluid ingress, and the observed short circuits.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the short circuits and leaks remain unknown.
|
|
Event Description
|
This report is to advise of multiple short circuits and leaks noted during analysis.
|
|