VENUSA DE MEXICO S.A. DE C.V. TACTICATH¿ QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number PN-004 075
Device Problems Fluid/Blood Leak (1250); Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/15/2022
Event Type  malfunction  
Manufacturer Narrative
One 75 mm tacticath quartz contact force ablation catheter was received for evaluation.Multiple short circuits were detected.The device did not meet specifications during a shaft leak test and an irrigation leak test.The stainless steel irrigation tubing was fractured and detached from the deformable body at the location of the fractured spine, consistent with the failed leak tests, fluid ingress, and the observed short circuits.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the short circuits and leaks remain unknown.
 
Event Description
This report is to advise of multiple short circuits and leaks noted during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
VENUSA DE MEXICO S.A. DE C.V.
calle hertz 1525-6
parque industrial j. bermudez
ciudad juarez, chihuahua 32470
MX  32470
Manufacturer (Section G)
VENUSA DE MEXICO S.A. DE C.V. (JUAREZ)
calle hertz 1525-6
parque industrial j. bermudez
ciudad juarez, chihuahua 32470
MX   32470
MDR Report Key15430867
MDR Text Key303905318
Report Number9680001-2022-00044
Device Sequence Number1
Product Code OAE
UDI-Device Identifier07640157990040
UDI-Public07640157990040
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P130026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Physician
Type of Report Initial
Report Date 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/25/2022
Device Model NumberPN-004 075
Device Catalogue NumberPN-004075
Device Lot Number7526048
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-