COCHLEAR LIMITED NUCLEUS CI622 COCHLEAR IMPLANT WITH SLIM STRAIGHT ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number CI622
Device Problem Malposition of Device (2616)
Patient Problems Ossification (1428); Fibrosis (3167)
Event Date 03/04/2022
Event Type  Injury  
Manufacturer Narrative
This report is submitted on september 16, 2022.
 
Event Description
Per the clinic, the patient experienced poor performance with the device.The device was explanted due to a coiling of the electrode in lump in the basal turn of the cochlea on (b)(6) 2022.Hypothesis of fibrosis with onset ossification around the electrode causing the drop in performance thus preventing reimplantation.
 
Manufacturer Narrative
Device analysis report is attached.This report is submitted on november 24, 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS CI622 COCHLEAR IMPLANT WITH SLIM STRAIGHT ELECTRODE
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LIMITED
1 university avenue
macquarie university
nsw 2109
AS  2109
Manufacturer (Section G)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS   2109
Manufacturer Contact
ken yian chow
unit ug-1, vertical podium
no. 8 jalan kerinchi,
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key15434062
MDR Text Key300022891
Report Number6000034-2022-02744
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09321502036573
UDI-Public(01)09321502036573(11)211122(17)231121
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date11/21/2023
Device Model NumberCI622
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/31/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-