Lot Number 0028955094
Device Problem Premature Activation (1484)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/04/2022
Event Type  Injury  
Event Description
It was reported that stent premature deployment occurred.The patient presented for percutaneous coronary intervention (pci).The stenosed target lesion was located in the right coronary artery (rca).A 4.00 x 20mm synergy xd drug-eluting stent was advanced for treatment.However, before reaching the correct location, it was noticed that the stent had flared on both proximal and distal ends.It was noted that without touching the inflation device, it was measuring 2 atmospheres already.Subsequently, the stent had to be deployed on the spot.A 4.00 x 24 mm synergy xd drug-eluting stent was used to fully cover the lesion and complete the procedure.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15438822
MDR Text Key300054995
Report Number2124215-2022-33391
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/15/2024
Device Lot Number0028955094
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/25/2022
Initial Date FDA Received09/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
INFLATION DEVICE: ENCORE 26
Patient Outcome(s) Required Intervention;
-
-