• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US DXTEND METAGLENE; SHOULDER METAGLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US DXTEND METAGLENE; SHOULDER METAGLENE Back to Search Results
Model Number 130760000
Device Problems Device Dislodged or Dislocated (2923); Osseointegration Problem (3003)
Patient Problem Inadequate Osseointegration (2646)
Event Date 09/02/2022
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).The device catalog number is unknown; therefore, udi is unavailable.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that screws broke off and metaglene with glenosphere detached.Surgeon removed broken bits and implanted competitor custom baseplate and humeral head with replacement depuy humeral cup poly.Broken pieces were retrieved and to be sent in.There was loosening reported at the bone to implant interface.Doi: unknown.Dor: (b)(6) 2022.Affected side: right shoulder.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : the device associated with this complaint was returned for examination.Visual examination found the device devoid of ha coating and any kind of debris, suggesting loosening of the device, confirming the reported allegation.The reported allegation for disassociation cannot be confirmed with the information provided.Depuy synthese considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the investigation was not able to retrieve the required lot code.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND METAGLENE
Type of Device
SHOULDER METAGLENE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15443131
MDR Text Key300122925
Report Number1818910-2022-18127
Device Sequence Number1
Product Code HSD
UDI-Device Identifier10603295027737
UDI-Public10603295027737
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 09/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/17/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number130760000
Device Catalogue Number130760000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/07/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DXTEND GLENOSPHERE STD D38MM; DXTEND METAGLENE; DXTEND STAND PE CUP D38 +3MM; UNKNOWN SHOULDER GLENOSPHERE; UNKNOWN SHOULDER LOCKING SCREW; UNKNOWN SHOULDER LOCKING SCREW; UNKNOWN SHOULDER NON-LOCKING SCREW; UNKNOWN SHOULDER NON-LOCKING SCREW
Patient Outcome(s) Required Intervention;
Patient Age67 YR
Patient SexFemale
-
-