SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C37101329-NLJ |
Device Problems
Leak/Splash (1354); Tear, Rip or Hole in Device Packaging (2385)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.The sample was visually inspected at a distance of 12" under normal lighting to received unit, in order to detect any damage on the unit.Embedded resin was found on the tube.Sample was tested.The reported issue was confirmed.A corrective and preventative action has been opened to determine the root cause and implement proper corrective actions.D4 udi is unknown.D5 operator of device is unknown.G5 pma/510k is unknown.No information has been provided to date.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# 617147.
|
|
Event Description
|
It was reported that during a pre-use check, an air leak from the product was detected.Also, a tear in the anesthesia circuit was found.No patient injury was reported.
|
|