MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. AMPERE¿ RF ABLATION GENERATOR; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. AMPERE¿ RF ABLATION GENERATOR; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Device Problem Impedance Problem (2950)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/01/2022
Event Type  malfunction  
Event Description
During the ablation procedure, signals could not be obtained and an impedance issue was noted resulting in procedure cancellation.It was not possible to get ablation signals on the systems.Additionally, the generator displayed an impedance out of range message with all 3 catheters.The issue was resolved by replacing the tacticath se catheter with no consequences to the patient.The signal issue with the ablation catheters was due to the gen connect cable being loose from the generator to x.However, after that issue was resolved, the generator read impedance out of range after 1-2 minutes with all 3 ablation catheters.The generator was shut down, turned back on and the impedance was normal for 1-2 minutes.After time spent troubleshooting cable issues, the patients pathway was non inducible, and the procedure was cancelled.
 
Manufacturer Narrative
The results of investigation are inconclusive as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported impedance issue and subsequent cancellation remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPERE¿ RF ABLATION GENERATOR
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15446444
MDR Text Key300578213
Report Number2184149-2022-00220
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/01/2022
Initial Date FDA Received09/19/2022
Supplement Dates Manufacturer Received10/10/2022
Supplement Dates FDA Received10/10/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age13 YR
Patient SexMale
-
-