Model Number 71992-01 |
Device Problems
Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
|
Patient Problems
Hypoglycemia (1912); Muscle Weakness (1967); Loss of consciousness (2418)
|
Event Date 09/06/2022 |
Event Type
Injury
|
Event Description
|
A complaint was received reporting a sensor error message while using the adc device.As a result, the customer experienced weakness and lost consciousness.The customer was provided a glucagon injection by a non-healthcare professional for treatment.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.No product has been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the libre sensor and libre sensor kits were reviewed and the dhrs showed the libre sensor and libre sensor kits passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A complaint was received reporting a sensor error message while using the adc device.As a result, the customer experienced weakness and lost consciousness.The customer was provided a glucagon injection by a non-healthcare professional for treatment.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
Sensor (b)(6) has been returned and investigated.The sensor plug is properly seated and no physical damage was observed.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 5 (indicating normal termination).Visual inspection has been performed on the sensor plug assembly, no failure modes were observed.Sim vivo testing (simulation of the electrical signal produced by the sensor tail) was performed and all results were within specification.No malfunction or product deficiency was identified.Therefore, the issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A complaint was received reporting a sensor error message while using the adc device.As a result, the customer experienced weakness and lost consciousness.The customer was provided a glucagon injection by a non-healthcare professional for treatment.There was no report of death or permanent injury associated with this event.
|
|
Search Alerts/Recalls
|