Model Number CI-1400-01 |
Device Problems
Material Protrusion/Extrusion (2979); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Unspecified Infection (1930)
|
Event Type
Injury
|
Event Description
|
The recipient reportedly experienced an infection and device extrusion.The recipient's device was explanted.The recipient was given antibiotics (type unknown), and is doing okay.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
The recipient's infection has reportedly resolved.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock was verified.The condition of the electrode prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed all of the tests performed.This version of the hires 90k device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|