• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM MEXICO ADVANTA 2 FRAME; BED, MANUAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM MEXICO ADVANTA 2 FRAME; BED, MANUAL Back to Search Results
Model Number P1190A000016
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/22/2022
Event Type  malfunction  
Manufacturer Narrative
The customer replaced the mcb, scale board, external alarm, and siderail interface board with no change.The hillrom technician suggested replacing the cable connecting the external alarm to the siderail interface board to resolve the reported event.Per the hillrom service manual, the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Brake alarm system: connect the bed to a wall power source, and set the brake to neutral.Make sure that the alarm activates and that you can hear it.If you cannot hear the alarm, troubleshoot the alarm and replace parts as necessary.Make sure that the alarm deactivates when the brake is set while the bed is connected to a wall power source.Make sure that the brake alarm switch is in the correct position.Adjust or replace parts as necessary.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician suggested replacing the cable connecting the external alarm to the siderail interface board to resolve the reported event.Hillrom technical support contacted the account two additional times trying to obtain the resolution for the reported event.No response has been received from the account.Based on this information, no further action is required.
 
Event Description
Hillrom received a report from a hillrom technician stating the bed brakes not set alarm was not working.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTA 2 FRAME
Type of Device
BED, MANUAL
Manufacturer (Section D)
HILL-ROM MEXICO
ave.del telefono no. 200 col. huinala
apodaca, nuevo leon 66640
MX  66640
Manufacturer Contact
brad wheeler
1069 state route 46 east
batesville, IN 47006
3128199307
MDR Report Key15455426
MDR Text Key302351834
Report Number3006697241-2022-00059
Device Sequence Number1
Product Code FNJ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP1190A000016
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/22/2022
Initial Date FDA Received09/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-