Manufacturer's investigation conclusion: the reported event of the centrimag system stopping atypically was not confirmed.The centrimag console (serial number (b)(4)) was not returned for analysis, and no log files were associated with the reported event.Questions regarding the event were asked; however, no additional information was able to be provided.This event will be reopened with further investigation if the console is returned for analysis.The root cause of the reported event was unable to be determined.Review of the device history record for centrimag 2nd gen.Primary console, serial number (b)(4), showed the device was manufactured in accordance with manufacturing and quality assurance specifications.The 2nd generation centrimag system operating manual provides information regarding emergencies/troubleshooting in section 9.The recommended practice whenever there is a console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the back-up motor and console.Switch all components (console, motor, flow probe and cables) simultaneously to continue patient support, and then perform troubleshooting on the non-functioning system, when it is no longer being used for patient support.No further information was provided.The manufacturer is closing the file on this event.
|