• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Catalog Number C45101316J
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/18/2020
Event Type  malfunction  
Manufacturer Narrative
A used sample was received to perform an investigation.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A visual inspection of the sample at a distance of 12 inches under normal lighting was performed and found no discrepancies.Functional testing was done using a leak test and it was found that the circuit failed the leak test.The root cause of the reported problem was unable to be determined.A corrective and preventive action (capa) was initiated to address this problem.No information has been provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event Description
It was reported that during a hospital pre-use check, air leaked from the breathing circuit component.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ANESTHESIA BREATHING CIRCUIT
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
aichi pref.
minneapolis, MN 55442
MDR Report Key15457038
MDR Text Key306018587
Report Number3012307300-2022-19824
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date10/03/2022
Device Catalogue NumberC45101316J
Device Lot Number3892065
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2020
Was the Report Sent to FDA? No
Date Manufacturer Received07/09/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/04/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-