NULL PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Device Problem
Defective Component (2292)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/01/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during a pre-use check, leakage of air from the product was observed.The customer suspected something was wrong with the breathing bag.No patient injury was reported.
|
|
Manufacturer Narrative
|
No lot number was provided; therefore, device history record review could not be completed.A product sample was received for evaluation.Visual and functional testing were performed.The sample was visually inspected at a distance of 12" under normal lighting to received unit, in order to detect any damaged components.No discrepancies detected.The sample passed the leak test.No root cause could be determined since the complaint was not confirmed.Udi is unknow.Operator of device is unknown.Manufacturing site address is unknown.Pma/510k is unknown.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
|