SMITH & NEPHEW, INC. JRNY II TKA FEM TRL IMPACTOR SZ1-10; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
Model Number 74012514 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/30/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that, during a tka surgery, the metal slot of the jrny ii tka fem trl impactor sz1-10 that holds trial in place broke off.No pieces broke inside patient.The procedure was completed, without delay, using the same device.Patient was not harmed.
|
|
Manufacturer Narrative
|
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirms the trial impactor head broke off.The trial impactor head was returned with the device.The device shows signs of significant wear and use.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device broke and it cannot longer fit its purpose, the contribution of the device to the reported event could be corroborated.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|