Model Number 1458Q/86
Device Problem Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2022
Event Type  malfunction  
Event Description
It was reported a patient presented for an implant procedure on (b)(6) 2022.During procedure the guidewire inserted into the left ventricular lead became stuck.The lead was replaced within the same operation.Post-procedure the patient was stable.
 
Manufacturer Narrative
The reported event of guidewire stuck inside the lead was confirmed.As received, a complete lead without the guidewire was returned in one piece for analysis.Visual inspection of the lead found the inner coil was stretched/damaged at the s-curve region.The guidewire insertion test couldn¿t perform due to the damaged inner coil.The cause of the reported event was isolated to the damaged inner coil.The s-curve hump height was measured within specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15472409
MDR Text Key304441092
Report Number2017865-2022-39129
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeCH
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000126360
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-