Model Number D97130F5 |
Device Problem
Pacing Problem (1439)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/05/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reported event of pacing issue was unable to be confirmed.No visible damage or abnormality was observed from catheter body, balloon, windings and returned syringe.Continuity testing was performed on the distal and proximal circuits and there were no open, intermittent, or short conditions observed.The balloon inflated clear and concentric with 1.3 cc air and the balloon remained inflated for 5 min.Without leakage.Further evaluation regarding related quality issues is under investigation.A device history record review was completed and documented that device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|
|
Event Description
|
It was reported a swan ganz catheter was unable to pace on the first day of use.Information such as the background of malfunction occurrence, the phase when the catheter got unable to pace, what kind of surgery or examination the catheter was used for, if the patient had cardiac conduction defect or if the catheter was replaced is unknown.Patient demographic information was requested but unavailable.There were no patient complications reported.
|
|