EDWARDS LIFESCIENCES PR SWAN-GANZ BIPOLAR PACING CATHETER; CATHETER, FLOW DIRECTED Back to Search Results
Model Number D97130F5
Device Problem Pacing Problem (1439)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/05/2022
Event Type  malfunction  
Manufacturer Narrative
The reported event of pacing issue was unable to be confirmed.No visible damage or abnormality was observed from catheter body, balloon, windings and returned syringe.Continuity testing was performed on the distal and proximal circuits and there were no open, intermittent, or short conditions observed.The balloon inflated clear and concentric with 1.3 cc air and the balloon remained inflated for 5 min.Without leakage.Further evaluation regarding related quality issues is under investigation.A device history record review was completed and documented that device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Event Description
It was reported a swan ganz catheter was unable to pace on the first day of use.Information such as the background of malfunction occurrence, the phase when the catheter got unable to pace, what kind of surgery or examination the catheter was used for, if the patient had cardiac conduction defect or if the catheter was replaced is unknown.Patient demographic information was requested but unavailable.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ BIPOLAR PACING CATHETER
Type of Device
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco 00610
*  00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco 00610
*   00610
Manufacturer Contact
jonathan diaz
1 edwards way
irvine, CA 92614
MDR Report Key15476723
MDR Text Key305883392
Report Number2015691-2022-08068
Device Sequence Number1
Product Code DYG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K822723
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date03/20/2024
Device Model NumberD97130F5
Device Catalogue NumberD97130F5
Device Lot Number64276258
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/05/2022
Initial Date FDA Received09/24/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/23/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-